Application of advanced high resolution mass spectrometric techniques for the analysis of losartan potassium regarding known and unknown impurities

Laura Backer,Helmut Buschmann,Martina Kinzig,Fritz Sörgel,Oliver Scherf-Clavel,Ulrike Holzgrabe
DOI: https://doi.org/10.1016/j.jpba.2023.115955
IF: 3.571
2024-01-06
Journal of Pharmaceutical and Biomedical Analysis
Abstract:Recalls of medicinal products can cause supply bottlenecks. This is often due to the findings of unexpected impurities that pose a health risk to patients. A recent example is losartan potassium which was contaminated with azido-impurities. The choice of the analytical method determines which substances can be detected and thus controlled. In this study a combination of an untargeted screening approach for impurities and a targeted evaluation of high-resolution mass spectrometry data was applied to search for impurities not described so far, leaving out a precise quantification. Six losartan potassium samples were studied regarding known and unknown impurities and hence highlight the applicability and capability of the approach. For separation a Zorbax RR StableBond C18 column (3.0 ×100 mm, particle size of 3.5 μm, pore size of 80 Å), a gradient elution and an electrospray ionization in positive and negative mode for mass spectrometric detection was used. An information-dependent acquisition method was applied for the measurement of losartan potassium samples. The untargeted data evaluation using general unknown comparative screening revealed the presence of N -methyl-2-pyrrolidone (NMP) and another impurity from synthesis. The identity of NMP was corroborated by a spiking experiment and the amount was estimated by means of standard addition. A targeted data evaluation by generating extracted ion chromatograms resulted in finding of four additional impurities. Combined approaches like this are needed to detect and respond to changes in the quality of drugs precociously.
pharmacology & pharmacy,chemistry, analytical
What problem does this paper attempt to address?