Development and Validation of Stability Indicating Chromatographic Methods for Determination of Azilsartan Medoxomil in Pharmaceutical Formation

Mukesh Kher,Vaibhav Bhatt,Aashka Jani,Navin Sheth
DOI: https://doi.org/10.1080/22297928.2020.1784788
2020-05-03
Analytical Chemistry Letters
Abstract:Two stability-indicating chromatographic methods namely, reverse-phase high-performance liquid chromatography (RP-HPLC) and high-performance thin-layer chromatography (HPTLC) were developed and validated for the determination of azilsartan medoxomil from pharmaceutical dosage form in presence of its degradation products as per ICH guideline. In RP-HPLC, separation was performed with Hiber® C-18 column (250 mm X 4.6 mm, 5 μm), using mobile phase methanol: acetonitrile: water (88:8:4 v/v/v) at a flow rate of 0.5 ml/min. The analyte was detected at 254 nm over a concentration range of 20 - 70 μg/ml with correlation coefficient of 0.999. In HPTLC, separation was carried out with silica gel G60 F254 aluminum sheet using acetone: toluene: ammonia (8.2:1.7:0.1 v/v/v) as a developing system. Linearity of the proposed HPTLC method was found 500 - 900 ng/band and correlation coefficient 0.9970 at 254 nm. Besides, stress degradation study was also carried out wherein degraded product peaks were well resolved from the analyte peak with difference in their retention time and retardation factor in RP-HPLC and HPTLC methods respectively. Statistical analysis proved that there were no significant differences between the two validated methods. The proposed stability-indicating methods can be applied for the analysis of azilsartan medoxomil in pharmaceutical formulations for quality control.
What problem does this paper attempt to address?