Enhancement of Convection and Molecular Transport into Film Stacked Structures by Introduction of Notch Shape for Micro-Immunoassay

Daiki Arai,Satoshi Ogata,Tetsuhide Shimizu,Ming Yang
DOI: https://doi.org/10.3390/mi15050613
IF: 3.4
2024-05-01
Micromachines
Abstract:A 3D-stack microfluidic device that can be used in combination with 96-well plates for micro-immunoassay was developed by the authors. ELISA for detecting IgA by the 3D-stack can be performed in one-ninth of the time of the conventional method by using only 96-well plates. In this study, a notched-shape film was designed and utilized for the 3D-stack to promote circulation by enhancing and utilizing the axial flow and circumferential flow in order to further reduce the reaction time. A finite element analysis was performed to evaluate the axial flow and circumferential flow while using the 3D-stack in a well and design the optimal shape. The 3D-stack with the notched-shape film was fabricated and utilized for the binding rate test of the antibody and antigen and ELISA. As a result, by promoting circulation using 3D-stack with notched-shape film, the reaction time for each process of ELISA was reduced to 1 min, which is 1/60 for 96 wells at low concentrations.
nanoscience & nanotechnology,instruments & instrumentation,physics, applied,chemistry, analytical
What problem does this paper attempt to address?