Optimising total knee replacement imaging: a novel 3D printed PET/CT anthropomorphic phantom for metal artefact simulation

Rajeh Assiri,Karen Knapp,Jon Fulford,Junning Chen
DOI: https://doi.org/10.1186/s40658-024-00634-2
2024-03-29
EJNMMI Physics
Abstract:Arthroplasty phantoms, including total knee replacement (TKR) phantoms, have been frequently used to test metal artefact reduction methods applied to positron emission tomography/computed tomography (PET/CT) images. These phantoms generally simulate either simple anatomical features or simple activity distribution around the metal inserts in the PET/CT scans. 3D printing has been used recently to fabricate fillable anthropomorphic phantoms that accurately simulate volume and geometry. This study aims to describe the process of image segmentation, phantom modelling, 3D printing and validation of a population-based fillable TKR phantom that simulates human TKR PET/CT metal artefacts.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?