Constant Fraction Discriminator for NA62 experiment at CERN

M. Zamkovsky,P. Lichard,the NA62 collaboration
DOI: https://doi.org/10.1088/1748-0221/19/03/c03002
2024-03-07
Journal of Instrumentation
Abstract:A new Constant Fraction Discriminator with additional Time over Threshold measurement capabilities will be presented. It operates in a wide dynamic range of 1:150, with an excellent time resolution of better than 70 ps over one order of magnitude of the input signals. It is highly customizable for different signal shapes and thresholds, using remotely-programmable parameters for Detector Control System commands. Two outputs, each in the Nuclear Instrumentation Module and Low Voltage Differential Signaling standards, provide a precise signal arrival time and Time over Threshold information with programmable thresholds. The technical specification and performance measured with cosmic rays and in the high-intensity NA62 experiment will be reported.
instruments & instrumentation
What problem does this paper attempt to address?