Design and Implementation of a Compact Analog Constant Fraction Discriminator for High-Resolution Timing in Gamma-Ray Spectroscopy

Michael Wiebusch
2024-11-01
Abstract:This work presents a custom analog front-end card designed for the read-out of PMTs coupled to lanthanum bromide scintillators. It integrates 16 discrete analog constant fraction discriminators (CFDs) on a compact 12x10 cm board, providing precise timing information for nuclear lifetime measurements. The design emphasizes cost-effectiveness, utilizing off-the-shelf discrete components, as well as compactness, achieved by using miniature coaxial connectors and cables as delay elements. The focus of this paper lies on the unusual and extremely minimalist analog shaper/discriminator design which is devised without operational amplifiers, making use of only a handful of RF transistors and LVDS receivers in place of comparators.
Instrumentation and Detectors,High Energy Physics - Experiment
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to achieve high - resolution time measurement in gamma - ray spectroscopy. Specifically, the author designed and implemented a compact analog constant - fraction discriminator (CFD) to improve the time precision in nuclear lifetime measurement. Traditional methods are affected by the so - called "time - drift" effect when dealing with fast signals, that is, the systematic deviation of the measured arrival time due to different signal amplitudes. The main purpose of CFD is to trigger based on a fixed proportion of the signal peak amplitude rather than on the leading edge of the signal, thus effectively offsetting this time - drift effect. The focus of the paper is on presenting a very minimalist analog shaping / discrimination circuit design without using operational amplifiers. This design uses a small number of RF transistors and LVDS receivers instead of comparators, which not only saves power consumption, cost and board space, but also maintains high performance. The designed CFD is particularly suitable for processing pulse signals generated by lanthanum bromide scintillator detectors coupled with photomultiplier tubes (PMT), but its concept can be applied to a wider range of detector types.