Timing performances of a Time-of-Flight detection system for the FRACAS large acceptance mass spectrometer

S. Salvador,E. Barlerin,J. Perronnel,C. Vandamme
DOI: https://doi.org/10.48550/arXiv.1911.13249
2020-02-12
Abstract:In this work, we report on the timing performances of the Time-Of-Flight (TOF) apparatus of the FRACAS large acceptance mass spectrometer designed to measure the fragmentation cross sections of a $^{12}$C beam in hadrontherapy. The TOF system is composed of a Parallel Plate Avalanche Counter and scintillating detectors to create so-called $\Delta$E---TOF maps used in the fragment charge identification process. From tests with an alpha source and a $^{12}$C beam experiment, we measured a coincidence resolving time for the apparatus below 300 ps. Those results should lead in the final design to the identification of all the fragments produced by the interaction of $^{12}$C ions with thin targets for energies up to 400 MeV/nucleon.
Instrumentation and Detectors,Nuclear Experiment
What problem does this paper attempt to address?