Abstract:We report on the use of an optically trapped microsphere as an acoustic transducer. A model for the hydrodynamic coupling between the microsphere and the surrounding acoustic fluid flow is combined with thermomechanical calibration of microsphere-position tracking to enable quantitative acoustic measurements. We describe our technique in detail, including the self-noise, the sensitivity, and the minimum detectable signals, using a model appropriate for both liquid and gas environments. We then test our approach in an air-based experiment and compare our measurements with two state-of-the-art commercially available acoustic sensors. Piezoelectrically driven bursts of pure tones and laser ablation provide two classes of test sounds. We find that accurate measurements with a bandwidth of 1 MHz are possible using our technique, improving by several orders of magnitude the bandwidth of previous flow measurements based on optically trapped microspheres. https://doi.org/10.1103/PhysRevApplied.21.014031 © 2024 American Physical Society