Suppressing Grating Lobes for Transcranial Focused Ultrasound System by Frequency-Modulated Excitation

Xiongfei Qu,Guofeng Shen,Nan Wu,Hao Wu,Shan Qiao,Enmin Wang,Yazhu Chen,Han Wang
DOI: https://doi.org/10.1109/tuffc.2020.3047664
2021-02-01
Abstract:Transcranial focused ultrasound is a novel noninvasive therapeutic modality for glioblastoma and other disorders of the brain. However, because the phase aberrations caused by the skull need to be corrected with computed tomography (CT) images, the transcranial transducer is tightly fixed on the patient's head to avoid any variation in the relative position, and the focus shifting relies mainly on the capacity for electronic beam steering. Due to the presence of grating lobes and the rapid degradation of the focus quality with increasing focus-shifting distance, transcranial focus-shifting sonication may damage healthy brain tissue unintentionally. To reduce the risks associated with transcranial focused ultrasound therapy, linear frequency-modulated (FM) excitation is proposed. The k-space corrected pseudospectral time domain (PSTD) and acoustic holography approach based on the Rayleigh integral are combined to calculate the distribution of the deposited acoustic power. The corresponding simulation was performed with axial/lateral focus shifting at different distances. The distributions of the deposited acoustic power show that linear FM excitation can effectively suppress undesired prefocal grating lobes without compromising focus quality.
engineering, electrical & electronic,acoustics
What problem does this paper attempt to address?