Effects of skull properties on long-pulsed transcranial focused ultrasound transmission

Han Li,Isla Barnard,Tyler Halliwell,Xinyu Zhang,Andreas Melzer,Zhihong Huang
2024-12-06
Abstract:Transcranial low-intensity focused ultrasound can deliver energy to the brain in a minimally invasive manner for neuromodulation applications. However, long-pulsed sonication through the skull introduces significant wave interactions, complicating precise energy delivery to the target. We present a comprehensive examination of intracranial acoustic fields generated by focused ultrasound transducers and assess the characteristics of cranial bone that affect acoustic transmission. Acoustic field maps were generated at 88 regions of interest across 10 historical and 2 Thiel-embalmed human skull specimens with sonication at frequencies of 220 kHz, 650 kHz, and 1000 kHz. The average peak pressure insertion loss was 3.1$\pm$1.4 dB, 9.0$\pm$1.7 dB, and 14.6$\pm$4.1 dB, and the average power insertion loss was 5.0$\pm$2.4 dB, 14.9$\pm$3.2 dB, and 24.1$\pm$6.3 dB, respectively. The effect of skull thickness, skull density ratio, and skull curvature on intracranial peak pressure and power was investigated and linear fits produced. Our results demonstrate that Thiel-embalmed samples fall within the confidence intervals of fits for historical samples. The effects of angulation and spacing between the transducer and the skull were also investigated. Results indicate that wave superposition resulting from skull and transducer spacing could lead to a 30-40% uncertainty in peak recorded intracranial pressure.
Medical Physics,Biological Physics