Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt $γ$-ray measurements

E. Testa,M. Bajard,M. Chevallier,D. Dauvergne,F. Le Foulher,J. -C. Poizat,C. Ray,M. Testa,N. Freud,J. -M. Létang,J.-C. Poizat,J.-M. Létang
DOI: https://doi.org/10.48550/arXiv.0809.0185
2008-09-01
Atomic Physics
Abstract:By means of a time-of-flight technique, we measured the longitudinal profile of prompt $\gamma$-rays emitted by 73 MeV/u $^{13}$C ions irradiating a PMMA target. This technique allowed us to minimize the shielding against neutrons and scattered $\gamma$-rays, and to correlate prompt gamma emission to the ion path. This correlation, together with a high counting rate, paves the way toward real-time monitoring of the longitudinal dose profile during ion therapy treatments. Moreover, the time correlation between the prompt gamma detection and the transverse position of the incident ions measured by a beam monitor can provide real-time 3D control of the irradiation.