Few-seconds range verification with short-lived positron emitters in carbon ion therapy

Caterina Cuccagna,Giuseppe Battistoni,Maria Giuseppina Bisogni,Piergiorgio Cerello,Alberto Del Guerra,Veronica Ferrero,Elisa Fiorina,Matteo Morrocchi,Francesco Pennazio,Roberto Sacchi,Ugo Amaldi
DOI: https://doi.org/10.1016/j.ejmp.2024.103209
IF: 3.119
2024-01-29
Physica Medica
Abstract:In-beam PET (Positron Emission Tomography) is one of the most precise techniques for in-vivo range monitoring in hadron therapy. Our objective was to demonstrate the feasibility of a short irradiation run for range verification before a carbon-ion treatment. To do so a PMMA target was irradiated with a 220 MeV/u carbon-ion beam and annihilation coincidences from short-lived positron emitters were acquired after irradiations lasting 0.6 s. The experiments were performed at the synchrotron-based facility CNAO ( Italian National Center of Oncological Hadrontherapy ) by using the INSIDE in-beam PET detector. The results show that, with 3·10 7 carbon ions, the reconstructed positron emitting nuclei distribution is in good agreement with the predictions of a detailed FLUKA Monte Carlo study. Moreover, the radio-nuclei production is sufficiently abundant to determine the average ion beam range with a σ of 1 mm with a 6 s measurement of the activity distribution. Since the data were acquired when the beam was off, the proposed rapid calibration method can be applied to hadron beams extracted from accelerators with very different time structures.
radiology, nuclear medicine & medical imaging