Method and equipment for estimation activity nuclides in samples of finite dimensions

A. Zhukouski,E. Levonyck,N. Sugai,H. Sugai,M. Mogi,A. Khruschinski,S. Kutsen
DOI: https://doi.org/10.1016/j.apradiso.2024.111226
IF: 1.787
2024-02-07
Applied Radiation and Isotopes
Abstract:This study presents a method of measuring the activity of a specific radionuclide 234m Pa in samples placed in bulk transport containers under changing background conditions. The method makes it possible to measure specific activity of 234 m Pa in objects without the need for sampling. The change in the effective sample volume limited by the surfaces of the containers is considered depending on the density of the measured material and the energy of gamma radiation of the radionuclide. The high sensitivity of scintillation detector, supplemented by adequate Monte Carlo simulation, allows spectrum measurements to be taken in a short time (less than an hour) with subsequent determination of specific activity. A comparison of measurement results and calculation of 234m Pa activity in samples with different densities and compositions using the proposed algorithm, and those obtained by an HPGe spectrometer, demonstrated the high efficiency of the proposed solution.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear
What problem does this paper attempt to address?