Validation of the validity of the new wrist-type fully automatic blood pressure monitor: DBP-2242 according to ISO 81060-2:2018+AMD 1:2020

Yuehong Jin,Xiaoming Bei,Ying Jin
DOI: https://doi.org/10.1097/mbp.0000000000000716
2024-07-16
Blood Pressure Monitoring
Abstract:Objective Evaluation of the accuracy of the wrist-type fully automatic blood pressure (BP) monitor: DBP-2242 in adolescent and adult populations according to ISO 81060-2 : 2018+Amd.1 : 2020. Methods BP measurements were taken from the subjects using the same-arm sequential method, and BP data measured by a mercury sphygmomanometer was used as the standard. Results This study analyzed 262 sets of data from 89 subjects. According to criterion 1, the mean difference between the SBP of the test and reference devices was −0.57 ± 7.31 mmHg, and the DBP was −2.27 ± 7.17 mmHg, which is in accordance with the requirements. According to criterion 2, the average difference between the SBP was −0.57 ± 6.25 mmHg and the DBP was −2.27 ± 5.99 mmHg, which is in accordance with the requirements. Conclusion Wrist-type fully automatic BP monitor: DBP-2242 complies with ISO 81060-2 : 2018+Amd.1 : 2020 and can be used for BP measurement in adolescent and adult populations.