Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device

Joon Ho Moon,Myung-Kyun Kang,Chang-Eun Choi,Jeonghee Min,Hae-Young Lee,Soo Lim
DOI: https://doi.org/10.1038/s41598-020-75892-y
IF: 4.6
2020-11-04
Scientific Reports
Abstract:Abstract Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1 ± 17.9 years). The BP was estimated using InBodyWATCH with an individualized estimation based on a neural network model. Three paired sets of BPs from the two devices were compared using correlation analysis and Bland–Altman plots ( n = 105 paired BP readings). The correlations for both systolic and diastolic BP (SBP and DBP) between the two devices were high ( r = 0.964 and 0.939, both P < 0.001). The mean difference was 2.2 ± 6.1 mmHg for SBP and −0.2 ± 4.2 mmHg for DBP; these were not significant ( P = 0.472 for SBP and P = 0.880 for DBP). The proportions of estimated SBP/DBP obtained from the InBodyWATCH within ± 5 mmHg of manual SBP/DBP were 71.4%/83.8%; within ± 10 mmHg they were 86.7%/98.1%; and within ± 15 mmHg they were 97.1%/99.0%. The estimated BP from this wearable cuff-less device correlated highly with the manual BP and showed good accuracy, suggesting its potential to be used in ambulatory BP monitoring.