Validation of the Raycome model M2 ambulatory blood pressure monitor in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018)

Shijie Yang,Zhanyang Zhou,Huanhuan Miao,Hongye Zhang,Qiong Zhou,Mei Zhai,Yuqing Zhang
DOI: https://doi.org/10.1097/mbp.0000000000000697
2024-02-23
Blood Pressure Monitoring
Abstract:Objective The aim of this study was to evaluate the accuracy of the Raycome model M2 oscillometric upper-arm blood pressure (BP) monitor developed for ambulatory BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise. Method Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in the general population using the same arm sequential BP measurement method. Three cuffs of the test device were used for arm circumference 18–22 cm (small), 22–32 cm (medium) and 32–42 cm (large). Results For the general validation study, 106 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.5 ± 6.2/−0.2 ± 5.1 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.23/4.50 mmHg (systolic/diastolic). In the ambulatory validation study ( N = 35), the mean difference was 0.4 ± 5.9/−1.1 ± 5.8 mmHg. The Raycome model M2 performed well against the standard in both the general and ambulatory validations and the Bland–Altman plots did not show any systematic variation in the error. Conclusion These data show that the Raycome model M2 monitor meets the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and in the ambulatory setting, indicating its suitability for measuring BP in the general population.
peripheral vascular disease
What problem does this paper attempt to address?