Stereotactic radiotherapy of intracranial tumor beds on a ring‐mounted Halcyon LINAC

Joshua Misa,Shane McCarthy,William St. Clair,Damodar Pokhrel
DOI: https://doi.org/10.1002/acm2.14281
2024-01-28
Journal of Applied Clinical Medical Physics
Abstract:Purpose This study sought to evaluate the feasibility and efficacy of the Halcyon Ring Delivery System (RDS) for delivering stereotactic radiotherapy (SRT) treatments for intracranial tumors beds. Methods Ten previously treated brain SRT patients for 30 Gy in five fractions with non‐coplanar HyperArc plans on TrueBeam (6MV‐FFF) were replanned on Halcyon (6MV‐FFF) using the same number of arcs and Eclipse's AcurosXB dose engine. Plan quality evaluation metrics per SRT protocol included: PTV coverage, GTV dose (minimum and mean), target conformity indices (CI), heterogeneity index (HI), gradient index (GI), maximum dose 2 cm away from the PTV (D2cm), and doses to organs‐at‐risk (OAR). Additionally, patient‐specific quality assurance (QA) results and beam‐on‐time (BOT) were analyzed. Results The Halcyon RDS provided highly conformal SRT plans for intracranial tumor beds with similar dose to target. When benchmarked against clinically delivered HyperArc plans, target coverage, CI(s) and HI were statistically similar. The Halcyon plans saw no statistical difference in maximum OAR doses to the brainstem, spinal cord, and cochlea. Due to the machine's coplanar geometry, the Halcyon plans showed a decrease in optic pathway dose (0.75 Gy vs. 2.08 Gy, p = 0.029). Overall, Halcyon's coplanar geometry resulted in a larger GI (3.33 vs. 2.72, p = 0.008) and a larger D2cm (39.59% vs. 29.07%, p
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?