Results of an interlaboratory comparison for characterization of Platinum nanoparticles using single-particle ICP-TOFMS
Lyndsey Hendriks,Sara Taskula,Bodo Hattendorf,Garret Bland,Eduardo Bolea-Fernandez,Frank Vanhaecke,Jingjing Wang,Mohammed Baalousha,Marcus von der Au,Björn Meermann,Timothy Ronald Holbrook,Stephan Wagner,Stasia Harycki,Alexander Gundlach-Graham,Frank von der Kammer,Robert Bruenjes,Jovanna Kocic,G Lowry
DOI: https://doi.org/10.1039/d3nr00435j
IF: 6.7
2023-06-17
Nanoscale
Abstract:This study describes an interlaboratory comparison (ILC) among nine (9) laboratories to evaluate and validate the standard operation procedure for single-particle (sp) ICP-TOFMS developed within the context of the Horizon 2020 project ACEnano. The ILC was based on the characterization of two different Pt nanoparticle (NP) suspensions in terms of particle mass, particle number concentration, and isotope ratio. The two Pt NPs suspensions were measured using icpTOF instruments (TOFWERK AG, Switzerland). Two Pt NPs samples were characterized and sized to 40.4 ± 7 nm and 58.8 ± 8 nm, respectively. Particle size results showed <16% relative standard deviation (RSD) among all participating labs and <4% RSD after excluding the two outliers. A good agreement was achieved between the different participating laboratories regarding particle mass, but the particle number concentration results were more scattered, with <53% RSD among all laboratories, which is consistent with results from previous ILC studies conducted using ICP-MS instrumentation equipped with a sequential mass spectrometer. Additionally, the capabilities of sp-ICP-TOFMS to determine masses on a particle basis are discussed with respect to the potential for particle density determination. Finally, because quasi-simultaneous multi-isotope and multi-element determinations are a strength of ICP-TOFMS instrumentation, the precision and trueness of isotope ratio determinations were assessed. The average of 1000 measured particles yielded a precision of below ±1% for intensity ratios of the most abundant Pt isotopes, i.e. 194Pt and 195Pt, while the accuracy of isotope ratios with the lower abundant isotopes was limited by counting statistics.
materials science, multidisciplinary,physics, applied,nanoscience & nanotechnology,chemistry