Data analysis for nanoparticles characterization with single particle inductively coupled plasma mass spectrometry: From microsecond to millisecond dwell times

Jinhui Liu,Xing Wei,Chengxin Wu,Lingna Zheng,Meng Wang,Mingli Chen,Jianhua Wang
DOI: https://doi.org/10.1016/j.aca.2023.341114
IF: 6.911
2023-03-19
Analytica Chimica Acta
Abstract:Single particle-inductively coupled plasma-mass spectrometry (SP-ICP-MS) has become a powerful technique for nanoparticles (NPs) characterization. However, the accuracy of NPs characterization by SP-ICP-MS is greatly affected by data acquisition rate and the way of data processing. For SP-ICP-MS analysis, ICP-MS instruments typically apply microsecond to millisecond dwell times (10 μs–10 ms). Considering the duration of one nanoparticle event in the detector is 0.4–0.9 ms, NPs will show different data forms with microsecond or millisecond dwell times. In this work, the effects of dwell times from microsecond to millisecond (50 μs, 100 μs, 1 ms and 5 ms) on the data forms are discussed in SP-ICP-MS analysis. The data analysis and data processing for different dwell times is discussed in detail, including the measurement of transport efficiency (TE), the distinction of signal and background, the evaluation of diameter limit of detection (LOD d ) and the quantification of mass, size and particle number concentration (PNC) of NPs. This work provides data support for the data processing process and aspects to consider in the characterization of NPs by SP-ICP-MS, which is expected to provide guidance and reference for researchers in SP-ICP-MS analysis.
chemistry, analytical