Wireless optically pumped magnetometer MEG

Hao Cheng,Kaiyan He,Congcong Li,Xiao Ma,Fufu Zheng,Wei Xu,Pan Liao,Rui Yang,Dongxu Li,Lang Qin,Shuai Na,Bingjiang Lyu,Jia-Hong Gao
DOI: https://doi.org/10.1016/j.neuroimage.2024.120864
IF: 5.7
2024-09-27
NeuroImage
Abstract:The current magnetoencephalography (MEG) systems, which rely on cables for control and signal transmission, do not fully realize the potential of wearable optically pumped magnetometers (OPM). This study presents a significant advancement in wireless OPM-MEG by reducing magnetization in the electronics and developing a tailored wireless communication protocol. Our protocol effectively eliminates electromagnetic interference, particularly in the critical frequency bands of MEG signals, and...
radiology, nuclear medicine & medical imaging,neurosciences,neuroimaging