Preoperative Prediction of Shock Impedance for Subcutaneous Implantable Cardioverter Defibrillator Using Chest Computed Tomography

Akira Kasagawa,Ikutaro Nakajima,Yui Nakayama,Daisuke Togashi,Kenichi Sasaki,Hisao Matsuda,Tomoo Harada,Yoshihiro J Akashi
DOI: https://doi.org/10.1253/circj.cj-23-0229
2024-02-02
Circulation Journal
Abstract:BACKGROUND: High shock impedance is associated with conversion failure among patients with subcutaneous implantable cardioverter defibrillators (S-ICD). Currently, there is no preoperative assessment method for predicting high shock impedance. This study examined the efficacy of chest computed tomography (CT) as a preoperative evaluation tool to assess the shock impedance of S-ICDs.Methods and Results: The amount of adipose tissue adjacent to the device and anteroposterior diameter at the basal heart region were measured preoperatively using chest CT. We examined the correlation between these measurements and shock impedance at the conversion test. We enrolled 43 patients with S-ICDs (mean [±SD] age 54±15 years; body mass index 23±4 kg/m<sup>2</sup>; PRAETORIAN score 30-270 points; amount of adipose tissue 1,250±716 cm<sup>3</sup>), who underwent intraoperative conversion tests by inducing ventricular fibrillation, which was terminated with a 65-J shock. A sufficient concordance correlation coefficient was observed between the shock impedance and the amount of adipose tissue (r=0.616, P&lt;0.01) and anteroposterior diameter (r=0.645, P&lt;0.01). In multiple regression analysis, the amount of adipose tissue (β=0.439, P=0.009) and anteroposterior diameter (β=0.344, P=0.038) were identified as independent predictive factors of shock impedance.CONCLUSIONS: The preoperative CT-measured amount of adipose tissue and basal heart anteroposterior diameter are independent predictors of shock impedance. These parameters may be more accurate in identifying higher shock impedance in patients with S-ICDs.
cardiac & cardiovascular systems
What problem does this paper attempt to address?