Microbubble-Facilitated Ultrasound Catheter Ablation Causes Microvascular Damage and Fibrosis

Babak Nazer,David Giraud,Yan Zhao,Yue Qi,O'Neil Mason,Peter D Jones,Chris J Diederich,Edward P Gerstenfeld,Jonathan R Lindner,Peter D. Jones,Chris J. Diederich,Edward P. Gerstenfeld,Jonathan R. Lindner
DOI: https://doi.org/10.1016/j.ultrasmedbio.2020.09.007
2021-01-01
Abstract:<p>High-intensity ultrasound (US) ablation produces deeper myocardial lesions than radiofrequency ablation. The presence of intravascular microbubble (MB) contrast agents enhances pulsed-wave US ablation <em>via</em> cavitation-related histotripsy, potentially facilitating ablation in persistently perfused/conducting myocardium. US ablation catheters were developed and tested in the presence of MBs using <em>ex vivo</em> and <em>in vivo</em> models. High-frame-rate videomicroscopy and US imaging of gel phantom models confirmed MB destruction by inertial cavitation. MB-facilitated US ablation in an <em>ex vivo</em> perfused myocardium model generated shallow (2 mm) lesions and, in an <em>in vivo</em> murine hindlimb model, reduced perfusion by 42% with perivascular hemorrhage and inflammation, but no myonecrosis.</p>