Beam dynamics framework incorporating acceleration to define the minimum aperture in two focusing schemes for proton radiotherapy linac

M. Southerby,R. Apsimon
DOI: https://doi.org/10.1103/physrevaccelbeams.27.064401
2024-06-05
Physical Review Accelerators and Beams
Abstract:In this paper, a self-consistent transverse beam dynamics framework is demonstrated that incorporates acceleration into the transverse beam dynamics studies for a proton linac machine. Two focusing schemes are developed and discussed: the FODO-like scheme and the minimum aperture scheme. The FODO-like scheme is a simple scheme, requiring only one quadrupole per cavity. The scheme is analytically solved to minimize the beam size at the cavity entrance/exit and ensures a constant beam size along the lattice, with respect to adiabatic damping due to longitudinally accelerating rf cavities. The minimum aperture scheme describes the regime that matches the beam ellipse to the acceptance ellipse of a cavity, allowing for the smallest possible aperture, for a given cavity length. A simple approximation of an rf cavity map is determined to allow changes in particle energy along a lattice, and acceleration is assumed only in the longitudinal direction. https://doi.org/10.1103/PhysRevAccelBeams.27.064401 Published by the American Physical Society under the terms of the Creative Commons Attribution 4.0 International license. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI. Published by the American Physical Society
physics, particles & fields, nuclear
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to define a transverse beam dynamics framework of the minimum aperture in the proton radiotherapy linear accelerator through combination with acceleration. Specifically, the authors developed two focusing schemes: the FODO - like scheme and the Minimum Aperture Scheme (MAS), in order to achieve effective control of the beam size, thereby reducing beam loss and increasing beam acceptance. ### Main problems 1. **Minimum aperture problem**: - How to minimize the beam aperture while ensuring the minimum size of the beam at the entrance and exit of the cavity, given the length of the cavity. - This involves how to match the beam ellipse with the cavity acceptance ellipse to achieve the smallest possible aperture. 2. **Combination of transverse beam dynamics and acceleration**: - How to establish a self - consistent transverse beam dynamics framework when considering longitudinal acceleration. - This includes how to maintain the transverse focusing of the beam during the acceleration process and how to deal with the adiabatic damping effect caused by longitudinal acceleration. ### Specific methods 1. **FODO - like scheme**: - This scheme forms a periodic FODO cell by placing a quadrupole magnet in front of each cavity. - Through analytical solution, the size of the beam at the entrance and exit of the cavity is minimized, and the beam size along the lattice is ensured to be constant. - This scheme is suitable for simple focusing requirements and can be achieved with only one quadrupole magnet. 2. **Minimum Aperture Scheme (MAS)**: - This scheme places multiple quadrupole magnets upstream of the cavity to match the beam ellipse with the cavity acceptance ellipse, thereby achieving the minimum aperture. - This scheme requires more quadrupole magnets, but can provide a higher proportion of the actual acceleration length (i.e., a higher "real - estate gradient") in some special cases. ### Mathematical models - **Twiss parameters**: Describe the distribution of the beam in phase space, including β - function, α - function and γ - function. - **Acceleration effect**: Consider the acceleration effect through the Lorentz - invariant normalized emittance to ensure the transverse focusing of the beam during the acceleration process. - **Transfer matrix**: Used to describe the transfer characteristics of the beam between different elements (such as quadrupole magnets, drift sections and cavities). ### Conclusions - This paper shows how to optimize the transverse beam dynamics in combination with acceleration in the proton linear accelerator and proposes two effective focusing schemes. - The FODO - like scheme is suitable for simple focusing requirements, while the MAS scheme is more effective in situations where the aperture needs to be minimized. Through these studies, the authors provide important theoretical basis and technical support for the design of proton radiotherapy linear accelerators.