Clinical evaluation of subtracted pointwise encoding time reduction with radial acquisition-based magnetic resonance angiography compared to 3D time-of-flight magnetic resonance angiography for improved flow dephasing at 3 Tesla

Qing Fu,Xiao-Yong Zhang,Xian-Bo Deng,Ding-Xi Liu,Xiao-yong Zhang,Xian-bo Deng,Ding-xi Liu
DOI: https://doi.org/10.1016/j.mri.2020.08.015
IF: 3.13
2020-11-01
Magnetic Resonance Imaging
Abstract:<h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Objective</h3><p>Flow dephasing artifacts within intracranial ICA have been problematic for 3D time-of-flight magnetic resonance angiography (3D-TOF-MRA). This study aimed to evaluate pointwise encoding time reduction with radial acquisition subtraction-based MR angiography (PETRA-MRA) for decreasing flow dephasing artifacts compared to 3D-TOF-MRA in intracranial segments of internal carotid artery (ICA) at 3 T.</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Methods</h3><p>Sixty healthy participants and seven patients with intracranial ICA aneurysms were enrolled to undergo 3D-TOF-MRA and PETRA-MRA. Two radiologists each evaluated the image quality of healthy participants using a 4-point scale (1: the best and 4: the worst). Quantitative analysis of the extent of homogeneity in signal intensity within the ICA and intracranial aneurysms was conducted using a parameter <em>d</em>: the higher the <em>d</em> value, the greater the signal homogeneity. Wilcoxon signed rank test, Chi-square test and the weighted kappa (κ) statistic were used for statistical analyses.</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Results</h3><p>The image quality of PETRA-MRA with an overall score of 1.35 ± 0.53 was significantly better than that obtained with 3D-TOF-MRA, with an overall score of 3.50 ± 0.62 (Z = -9.56, <em>p</em> &lt; 0.001). The parameter <em>d</em> of PETRA-MRA was higher than that of 3D-TOF-MRA for both 60 healthy participants (0.97 ± 0.05, 0.87 ± 0.11; z = −13.21, p &lt; 0.001) and 7 patients with intracranial aneurysms (0.81 ± 0.18, 0.74 ± 0.16; z = −2.37, <em>p</em> = 0.018).</p><h3 class="u-h4 u-margin-m-top u-margin-xs-bottom">Conclusion</h3><p>Compared with conventional 3D-TOF-MRA, PETRA-MRA remarkably improved the image quality with reduced flow dephasing artifacts in segments of intracranial ICA.</p>
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?