Quantitative Flow Measurement after Placing a Flow Diverter for a Distal Internal Carotid Artery Aneurysm
Chien-Wei Chen,Ho-Fai Wong,Yu-Ling Ye,Yao-Liang Chen,Wei-Liang Chen,Chang-Hsien Ou,Yuan-Hsiung Tsai
DOI: https://doi.org/10.1136/neurintsurg-2016-012730
IF: 4.8
2016-01-01
Journal of NeuroInterventional Surgery
Abstract:ObjectivesTo evaluate the differences in arterial flow after flow diverter placement using quantitative flow measurements based on digital subtraction angiography (DSA).MethodsBetween November 2013 and November 2015, all patients who had flow diverters placed for distal internal carotid artery (ICA) aneurysms were reviewed. Patients in whom the stent was placed across the ostia of the ophthalmic artery (OphA) and anterior choroidal artery (AChA) were enrolled. Five regions of interest were selected: the proximal ICA (as a reference), terminal ICA, middle cerebral artery (MCA), anterior cerebral artery (ACA), OphA, and AChA. The values of the peak, time-to-peak (TTP), and area under the curve (AUC) were analyzed using a quantitative DSA technique.ResultsThe study enrolled 13 patients. The quantitative flow analysis showed improved flow in the terminal ICA (peak and AUC, p=0.036 and p=0.04, respectively), MCA (AUC, p=0.023), and ACA (AUC, p=0.006), and decreased flow in the OphA (peak and AUC, p=0.013 and p=0.005, respectively) and AChA (peak and subtracted TTP, p=0.023 and p=0.050, respectively) after flow diverter placement. Larger aneurysm volume was significantly correlated with decreased OphA flow after the procedure (peak and AUC, p=0.049 and p=0.037, respectively). Larger aneurysm volume also had a marginal correlation with increased distal ICA flow after the procedure, but this did not reach significance (peak and AUC, p=0.195 and p=0.060, respectively).ConclusionsWithout using extra contrast medium or radiation dosages, color-coded DSA enables quantitative monitoring of the cerebral circulation after flow-diverting treatment.