3D printed two-in-one on-capillary detector: Combining contactless conductometric and photometric detection for capillary electrophoresis

Liye Yang,Guangchao Pan,Piwang Zhang,Qiang Liu,Xing Liu,Yan Li,Ying Liang,Min Zhang
DOI: https://doi.org/10.1016/j.aca.2021.338427
IF: 6.911
2021-05-01
Analytica Chimica Acta
Abstract:<p>In this work, for the first time, a 3D printed two-in-one on-capillary detector, combining contactless conductometric detection (C<sup>4</sup>D) and photometric detection (PD), is fabricated for capillary electrophoresis (CE). The C<sup>4</sup>D Faraday shield (FS) is printed using electrically conductive composite polylactic acid (PLA) to minimize the stray capacitance. Non-conductive PLA is also used to print the insulator of FS to prevent the electrical conduction with two stainless steel electrodes. A novel collimator, consisting of two partially aligned pinholes, is printed by conductive material to collimate the light-emitting diode beam. The C<sup>4</sup>D detection has a signal-to-noise ratio of 1092 ± 2 for 200 μM potassium on a 25 μm id capillary. The PD detection shows excellent linearity with stray light down to 8% and an effective path length at 73% of a 75 μm id capillary. The analytical performance is demonstrated by CE separation and detection of cations. PD shows limits of detection (LODs) of 1.3, 0.9, and 1.7 μM for cobalt, copper and zinc, which are complexed with 4-(2-Pyridylazo) resorcinol, while C<sup>4</sup>D shows LODs of 1.2, 1.4, 21 and 2.6 μM for potassium, sodium, cobalt and zinc, respectively.</p>
chemistry, analytical
What problem does this paper attempt to address?