Accuracy Investigation of High-Frequency Core Loss Measurement for Low-Permeability Magnetic Materials

Yuki Sato,Yuji Uehara,Satoshi Okamoto,Shigeyoshi Yoshida,Yasushi Endo,Nobuhisa Ono,Nobuto Misono,Hirokazu Matsumoto
DOI: https://doi.org/10.1109/tmag.2023.3283955
IF: 1.848
2023-01-01
IEEE Transactions on Magnetics
Abstract:This study investigates the accurate measurements of high-frequency core loss using a capacitive cancellation method for low-permeability magnetic material, which leads to high leakage flux during measurement. We observe that the core loss is underestimated with the lower coupling of the core-under-test. To solve the issue, we propose a new correction method in which the effective permeability observed during the measurement is measured, and the core loss is corrected with that permeability. The core loss is corrected by the proposed method even when high leakage flux exists.
engineering, electrical & electronic,physics, applied
What problem does this paper attempt to address?