Fabrication and application of a continuously regenerated cationic impurity removal device for ion chromatography

Yihua Li,Tianyang Zhang,Feifang Zhang,Weiqing Chen,Bingcheng Yang
DOI: https://doi.org/10.1016/j.chroma.2024.464904
IF: 4.601
2024-04-12
Journal of Chromatography A
Abstract:A continuously regenerated cationic impurity removal device (CR-CRD) has been fabricated and applied for ion chromatography (IC). The removal of cationic impurities is realized by electrodialytically replacing the cationic impurities with hydronium ions. The device is configured in a sandwich structure and the central eluent channel is respectively isolated from both electrodes by stacked cation exchange membranes and a bipolar membrane (BPM) plus stacked anion exchange membranes. The eluent channel is packed with cation exchange resins in hydronium form and their continuous regeneration can be achieved by electrodialysis. A desirable feature of the device is gas-free, and no degasser is required. It showed sufficient ability to remove cationic impurities, as indicated by >99.9% removal of 10 mL of 1 mM LiOH solution injected (∼10 μmol) or continuous removal of 1 mM LiOH solution at the flow rate of 1 mL/min (1 μmol/min). A useful application was for sample pretreatment in nuclear power industry, by eliminating strong matrix interference of the sample containing LiOH (1 mM) and boric acid (2000 mg/L) with trace anion analysis.