Investigation on Parylene-C based neural electrodes by accelerated life test and reliability improvement using polyimide flexible cable

Ju-Hwan Kim,Dong-Hyun Baek,Dae Hwan Kim,Dong-Wook Park
DOI: https://doi.org/10.1016/j.cap.2022.05.001
IF: 2.856
2022-07-01
Current Applied Physics
Abstract:Since the encapsulation layer of the neural electrode plays an important role in determining the reliability of the device, more studies on the long-term reliability of Parylene-C are required. In this paper, we described the long-term reliability of Parylene-C based neural electrodes through an accelerated lift test and proposed a method to improve the reliability using polyimide flexible cable (PFC) attached to the Parylene-C device. The accelerated lifetime test was performed in phosphate-buffered saline at 50 °C. The electrochemical characteristics were acquired and analyzed to evaluate the reliability of the Parylene-C coated neural electrode for up to 56 days, which is equivalent to 115 days as an accelerated time. The proposed PFC method showed improved long-term stability of the device, demonstrating an increased lifetime, showing that testing conditions and device setup affect overall results. This study will help make reliable Parylene-C based neural electrode systems with improved moisture protection.
physics, applied,materials science, multidisciplinary
What problem does this paper attempt to address?