Application of Printed Circuit Board Technology to FT-ICR MS Analyzer Cell Construction and Prototyping

Franklin E. Leach,Randolph Norheim,Gordon Anderson,Ljiljana Pasa-Tolic
DOI: https://doi.org/10.1007/s13361-014-0952-6
IF: 3.262
2014-07-24
Journal of the American Society for Mass Spectrometry
Abstract:Although Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) remains the mass spectrometry platform that provides the highest levels of performance for mass accuracy and resolving power, there is room for improvement in analyzer cell design as the ideal quadrupolar trapping potential has yet to be generated for a broadband MS experiment. To this end, analyzer cell designs have improved since the field’s inception, yet few research groups participate in this area because of the high cost of instrumentation efforts. As a step towards reducing this barrier to participation and allowing for more designs to be physically tested, we introduce a method of FT-ICR analyzer cell prototyping utilizing printed circuit boards at modest vacuum conditions. This method allows for inexpensive devices to be readily fabricated and tested over short intervals and should open the field to laboratories lacking or unable to access high performance machine shop facilities because of the required financial investment.Figureᅟ
chemistry, physical,spectroscopy, analytical,biochemical research methods
What problem does this paper attempt to address?