Research of Unidirectional Ion Ejection in Printed-Circuit-Board Ion Trap

Zhang Zai-Yue,Yuan Guang-Zhou,He Yang,Qian Jie,Zhang Shu-Guang,Yao Ru-Jiao,Ding Chuan-Fan,Li Xiao-Xu
DOI: https://doi.org/10.1016/s1872-2040(17)60995-2
2017-01-01
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY (CHINESE VERSION)
Abstract:Printed-circuit-board ion trap (PCBIT) is a novel ion trap mass analyzer, which is capable of optimizing its internal electric field distributions to improve the analytical performance by adjusting the radio frequency (RF) voltage-divided ratio. This work introduced odd electric field components into the trapping volume to achieve unidirectional ion ejection by applying asymmetric RF voltages to x electrode pairs of PCBIT. In this case, the center of ion vibration was displaced away from the geometrical center of PCBIT and ions were ejected predominantly through one of x electrode pairs. The relationship between asymmetric voltage-divided ratio ΔV and internal electric field distributions was investigated by simulation software SIMION and AXSIM. In addition, the ion trajectories and simulated mass spectrum peaks were calculated. The results showed that, for ions with m/z 609 Th, a mass resolution of over 2500 and an ion unidirectional ejection efficiency of over 90% were achieved in PCBIT with ΔV = 20% at an appropriate frequency of AC. Using this method, the ion detection efficiency of PCBIT was significantly improved while maintaining a high mass resolution, which made the PCBIT more suitable for developing miniaturized mass spectrometer.
What problem does this paper attempt to address?