1950-nm diode laser-assisted microanastomoses (LAMA): an innovative surgical tool for hand surgery emergencies

Franck Marie Leclère,Michel Schoofs,Peter Vogt,Vincent Casoli,Serge Mordon
DOI: https://doi.org/10.1007/s10103-015-1711-z
2015-02-21
Lasers in Medical Science
Abstract:Based on previous observations, the 1950-nm diode laser seems to be an ideal wavelength for laser microvascular anastomoses. The data presented here, part of a larger ongoing study, assess its use in emergency hand surgery. Between 2011 and 2014, 11 patients were operated on for hand trauma with laser-assisted microanastomoses (LAMA) and prospectively analysed. LAMA was performed with a 1950-nm diode laser after placement of equidistant stitches. For vessel size <1.5 mm, the following laser parameters were used: spot size 400 μm, five spots for each wall, power 125 mW, and arterial/venous fluence 100/90 J/cm2 (spot duration 1/0.9 s). Mean operating time for arterial and venous microanastomoses was 7.3 ± 1.4 and 8.7 ± 1.0 min, respectively. Three anastomoses required a secondary laser application. Arterial and venous patency rates were 100 % at the time of surgery. The success rate for the 11 procedures assessed clinically and with the Doppler was 100 %. The technique is compared to the current literature. The 1950-nm LAMA is a reliable tool with excellent results in emergency hand surgery. The system is very compact and transportable for utilization in the emergency operating room.
engineering, biomedical,surgery
What problem does this paper attempt to address?