A Time Difference Method for Measurement of Phase Shift between Distributed Feedback Laser Diode (DFB-LD) Output Wavelength and Intensity

Yongning Liu,Jun Chang,Jie Lian,Zhaojun Liu,Qiang Wang,Cunguang Zhu
DOI: https://doi.org/10.3390/s150716153
IF: 3.9
2015-07-06
Sensors
Abstract:A time difference method to conveniently measure the phase shift between output wavelength and intensity of distributed feedback laser diodes (DFB-LDs) was proposed. This approach takes advantage of asymmetric absorption positions at the same wavelength during wavelength increase and decrease tuning processes in the intensity-time curve by current modulation. For its practical implementation, a measurement example of phase shift was demonstrated by measuring a time difference between the first time and the second time attendances of the same gas absorption line in the intensity-time curve during one sine or triangle modulation circle. The phase shifts at modulation frequencies ranging from 50 Hz to 50 kHz were measured with a resolution of 0.001π. As the modulation frequency increased the shift value increased with a slowed growth rate.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?