Noninvasive Respiration Monitoring of Different Sleeping Postures Using an RF Sensor

Nuerzati Resuli,Marjorie Skubic,Jung Myungki
DOI: https://doi.org/10.1109/bibm52615.2021.9669669
2021-12-09
Abstract:In this paper, we examine the accuracy of a respiration rate algorithm using a radio frequency (RF) sensor while testing different sleeping postures. The goal is to use the algorithm in a contactless monitoring system for psychiatric patients in an unstructured hospital setting, in which the sleeping posture is unknown. Here, we first explore different postures in a lab setting. The application in the psychiatric hospital requires a non-wearable, non-accessible sensor that can track patients’ motion in the bed, estimate the patient’s respiration rate during sleep, and estimate the patient’s restless time during the night. The RF signal was collected for thirteen different sleeping postures, using a Vayyar Radar system with a carrier frequency of 6.014 GHz to capture all reflections by the FMCW (frequency modulated continuous wave) signal. Respiration belt data are used as ground truth. Preliminary results indicate that RF sensors can capture the respiration torso movement in different sleeping postures at a distance of 2.3m. The respiration rate estimation is 90% accurate when the torso area is directed toward the RF sensor, 87% accurate when the head is reversed to the opposite side of bed, and 86% accurate when sitting in bed reading.
What problem does this paper attempt to address?