Research on the Wear Performance of Artificial Knee Joint with Different Geometric Dimension

Xian Wu,Mingyang Du,Jianyun Shen,Da Zeng,Bin Dai,Yuan Li,Hairong Wu
DOI: https://doi.org/10.1007/s40846-022-00679-7
IF: 2
2022-01-15
Journal of Medical and Biological Engineering
Abstract:PurposeThis study aims at investigating the wear behavior of ultrahigh molecular weight polyethylene (UHMWPE), which limits the longevity of total knee replacement surgery.MethodsThe tensile and friction tests were conducted on UHMWPE material to obtain material properties. The finite element simulation model of artificial knee joint was established. Then, the finite element simulation was performed on the artificial knee joint with different geometric dimensions.ResultsThe wear rate results from FEM simulation showed 1.8% error compared with physical vitro wear test. The wear map distribution obtained from the FEM simulation was completely located in that results from the physical vitro wear test.ConclusionsThe contact stress and area between tibial insert and femoral condyle increased with an increase in geometric dimension. Compared with the small geometric dimension, the large geometric dimension shows an increase of 17% in wear weight, but a decrease of 14% in wear depth. The wear area and depth on the left half side are less than right half side.
engineering, biomedical
What problem does this paper attempt to address?