Ultraflexible organic amplifier with biocompatible gel electrodes

Tsuyoshi Sekitani,Tomoyuki Yokota,Kazunori Kuribara,Martin Kaltenbrunner,Takanori Fukushima,Yusuke Inoue,Masaki Sekino,Takashi Isoyama,Yusuke Abe,Hiroshi Onodera,Takao Someya
DOI: https://doi.org/10.1038/ncomms11425
IF: 16.6
2016-04-29
Nature Communications
Abstract:In vivo electronic monitoring systems are promising technology to obtain biosignals with high spatiotemporal resolution and sensitivity. Here we demonstrate the fabrication of a biocompatible highly conductive gel composite comprising multi-walled carbon nanotube-dispersed sheet with an aqueous hydrogel. This gel composite exhibits admittance of 100 mS cm−2 and maintains high admittance even in a low-frequency range. On implantation into a living hypodermal tissue for 4 weeks, it showed a small foreign-body reaction compared with widely used metal electrodes. Capitalizing on the multi-functional gel composite, we fabricated an ultrathin and mechanically flexible organic active matrix amplifier on a 1.2-μm-thick polyethylene-naphthalate film to amplify (amplification factor: ∼200) weak biosignals. The composite was integrated to the amplifier to realize a direct lead epicardial electrocardiography that is easily spread over an uneven heart tissue.
multidisciplinary sciences
What problem does this paper attempt to address?