Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study

Robbert W. van Hamersvelt,Arnold M. R. Schilham,Klaus Engelke,Annemarie M. den Harder,Bart de Keizer,Harald J. Verhaar,Tim Leiner,Pim A. de Jong,Martin J. Willemink
DOI: https://doi.org/10.1007/s00330-017-4801-4
IF: 7.034
2017-04-03
European Radiology
Abstract:<h3 class="a-plus-plus">Abstract</h3> <span class="a-plus-plus abstract-section id-a-sec1"> <h3 class="a-plus-plus">Objectives</h3> <p class="a-plus-plus">To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols.</p> </span> <span class="a-plus-plus abstract-section id-a-sec2"> <h3 class="a-plus-plus">Methods</h3> <p class="a-plus-plus">Two validated anthropomorphic phantoms containing inserts of 50–200 mg/cm<sup class="a-plus-plus">3</sup> calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90–200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA).</p> </span> <span class="a-plus-plus abstract-section id-a-sec3"> <h3 class="a-plus-plus">Results</h3> <p class="a-plus-plus">At both 120 and 140 kVp, excellent correlations (R = 0.97, P &lt; 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm<sup class="a-plus-plus">3</sup> (-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm<sup class="a-plus-plus">3</sup> (-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R<sup class="a-plus-plus">2</sup> ≥ 0.970, P &lt; 0.001) with DXA were found.</p> </span> <span class="a-plus-plus abstract-section id-a-sec4"> <h3 class="a-plus-plus">Conclusions</h3> <p class="a-plus-plus">SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings.</p> </span> <span class="a-plus-plus abstract-section id-a-sec5"> <h3 class="a-plus-plus"> <em class="a-plus-plus">Key points</em> </h3> <p class="a-plus-plus">• <em class="a-plus-plus">Dual</em>-<em class="a-plus-plus">layer spectral detector CT allows for accurate bone mineral density quantification</em>.</p> <p class="a-plus-plus">• <em class="a-plus-plus">BMD measurements on SDCT are strongly linearly correlated to DXA</em>.</p> <p class="a-plus-plus">• <em class="a-plus-plus">SDCT</em>, <em class="a-plus-plus">acquired for several indications</em>, <em class="a-plus-plus">may allow for evaluation of osteoporosis</em>.</p> <p class="a-plus-plus">• <em class="a-plus-plus">This potentially opens up the possibility for opportunistic osteoporosis screening</em>.</p> </span>
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?