Experimental Demonstration of a 10 K Sorption Cryocooler Stage

J. J. Wu,S. Bard,W. Boulter,J. Rodriguez,R. Longsworth
DOI: https://doi.org/10.1007/978-1-4615-2522-6_185
1994-01-01
Abstract:The basic feasibility of a periodic quick-cooldown 10 K sorption cryocooler was demonstrated in laboratory experiments. This paper summarizes the experiment hardware design and test results. The zirconium-nickel hydride sorption pump demonstrated the high pumping speeds and low pressures (below 1.3 × 10-4 MPa) required to solidify and evaporatively cool the hydrogen refrigerant to an ultimate low temperature of 9.2 K at the simulated detector interface. Data from pump characterization tests established the empirical relation between the pressure (P, Pa) and pumping speed (m, standard liter per minute) as P = 57.45 m. Two separate cryostats, utilizing different phase separation techniques were built and tested. Both cryostats demonstrated the ability to cooldown in under 2 minutes and maintain a simulated detector heat load of 150 mW below 11 K for over 30 minutes. The wicks in each cryostat demonstrated the ability to retain liquid hydrogen in tests conducted in an upside-down adverse-gravity orientation. A sorption cryocooler utilizing the technology described herein is ideal for spacecraft sensor cooling applications that require periodic quick-cooldown operation, negligible vibration, low power consumption, and long-life (≥ 10 years).