Transcutaneous Energy Transmission System for a Totally Implantable Artificial Heart Using a Two-Wire Archimedean Spiral Coil

Tomoki Okinaga,Takahiko Yamamoto,Kohji Koshiji
DOI: https://doi.org/10.1109/embc46164.2021.9630900
2021-11-01
Abstract:This paper describes the application of a proposed spiral coil to the transformer of a transcutaneous energy transfer system for a totally implantable artificial heart. To reduce the number of rectifier components in the power receiving circuit, the shape of the power receiving transformer was reviewed. The results indicated that the power transmission efficiency between the transformers was almost the same as that of the receiving transformer with the same shape. In addition, the calculations indicated that the power transmission efficiency including that of the power receiving circuit was increased, and the number of components in the power receiving circuit was reduced.