Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation

Dae-Young Park,Dong Hui Lim,Sungsoon Hwang,Joo Hyun,Tae-Young Chung
DOI: https://doi.org/10.1186/s12886-017-0550-z
IF: 2.086
2017-08-24
BMC Ophthalmology
Abstract:BackgroundTo evaluate and compare the astigmatism prediction errors taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens (IOL) implantation.MethodsPhacoemulsification with toric Precizon IOL implantation was performed in 41 eyes with corneal astigmatism (range, 1 to 5 diopters (D)) determined by IOLMaster and SimK on Pentacam. Preoperative corneal astigmatism measurements were obtained from IOLMaster readings (IOLMaster, Baylor-IOLMaster, and Barrett-IOLMaster) and Pentacam readings (SimK, Baylor-SimK, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector derived by manual vector summation using corneal front and back astigmatism). Prediction error and intraclass correlation coefficient (ICC) between the measured (or calculated) astigmatism by IOLMaster and Pentacam and the estimated corneal astigmatism estimated by IOL toricity power and residual astigmatism were determined.ResultsThe centroid errors in prediction error with IOLMaster, SimK, Baylor-IOLMaster, Baylor-SimK, Barrett-IOLMaster, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector were 0.59@103, 0.61 @103, 0.37@161, 0.41@162, 0.24@171, 0.36@162, 0.42@106, 0.04@8, 0.07@82, and 0.03@82, respectively, in with-the-rule (WTR) astigmatism eyes at postoperative 3-month. They were 0.22@87, 0.20@74, 0.16@21, 0.54@10, 0.43@3, 0.33@19, 0.51@25, 0.31@58, 0.29@50, and 0.14@50 in against-the-rule (ATR) astigmatism eyes. Of the ten modalities, vector showed the lowest WTR astigmatism prediction error and the highest ICC between the predicted and the estimated corneal astigmatism for both WTR and ATR eyes.ConclusionVector summation using anterior and posterior corneal surface power taken with the Pentacam yields the least astigmatism prediction error and is a promising tool for determining toric IOL cylinder power.
ophthalmology
What problem does this paper attempt to address?