Performance Survey and Comparison Between Rapid Sterility Testing Method and Pharmacopoeia Sterility Test

Adriana Bugno,Deborah Pita Sanches Saes,Adriana Aparecida Buzzo Almodovar,Kamal Dua,Rajendra Awasthi,Daniela Dal Molim Ghisleni,Marici Tiomi Hirota,Wesley Anderson de Oliveira,Terezinha de Jesus Andreoli Pinto
DOI: https://doi.org/10.1007/s12247-017-9303-z
2017-12-05
Journal of Pharmaceutical Innovation
Abstract:The sterility test described in pharmacopoeial compendia requires a 14-day incubation period to obtain a valid analytical result. Therefore, the use of alternative methods to evaluate the sterility of pharmaceuticals, such as the BacT/Alert® 3D system, is particularly interesting, because it allows a reduced incubation period and lower associated costs. Considering that the BacT/Alert® 3D system offers several culture media formulations developed for this microbial detection system, the present study was aimed to evaluate and compare the performance of BacT/Alert® 3D with the pharmacopoeial sterility test. There was no significant difference between the ability of the culture media to allow detection of microbial contamination. However, the rapid sterility testing method allowed a more rapid detection of the challenge microorganisms, which indicates that the system is a viable alternative for assessing the sterility of injectable products.
pharmacology & pharmacy
What problem does this paper attempt to address?