Characterization of the TSI model 3086 differential mobility analyzer for classifying aerosols down to 1 nm

Mark R. Stolzenburg,Jacob H. T. Scheckman,Michel Attoui,Hee-Siew Han,Peter H. McMurry
DOI: https://doi.org/10.1080/02786826.2018.1456649
2018-04-19
Aerosol Science and Technology
Abstract:Measurement systems for particle sizing starting at 1 nm are used to bridge the gap between mass spectrometer measurements and traditional aerosol sizing methods, and thus to enable measurement of the complete size distribution from molecules and clusters to large particles. Such a measurement can be made using a scanning mobility particle sizer equipped with a diethylene glycol growth engine (e.g., TSI Model 3777 Nano Enhancer) along with a condensation particle counter, and a differential mobility analyzer (DMA) appropriate for such small sizes. Previous researchers have used high-resolution DMA (HRDMA) and also the TSI Nano-DMA (Model 3085) in such a scanning mobility particle sizer system. In this study, we evaluate the performance of the recently introduced TSI 1 nm-DMA (Model 3086). The transfer function was characterized using 1–2 nm monomobile molecular ion standards. The same measurements were repeated on a TSI Nano-DMA, with good agreement to previously published values. From the measured transfer function, the resolution of each DMA model was determined as a function of particle size and sheath flow rate. Resolution of the TSI 3086 in the 1–2 nm range was 10–25% higher than the TSI 3085. Measured resolutions of the TSI 3086 were 10–20% lower than theoretically predicted values, whereas those of the Model 3085 were 0–10% lower. Copyright © 2018 TSI Inc.
environmental sciences,meteorology & atmospheric sciences,engineering, chemical, mechanical
What problem does this paper attempt to address?