Contribution of the Radial Distribution of the Flow Velocity to Band Broadening in HPLC Columns

Tivadar Farkas,Georges Guiochon
DOI: https://doi.org/10.1021/ac970530m
IF: 7.4
1997-11-01
Analytical Chemistry
Abstract:The radial distributions of the local linear velocity of the mobile phase, the local efficiency, and the local analyte concentration were determined at the outlet of 0.3−0.4 in. i.d. chromatographic columns by simultaneously recording the elution bands at different radial locations over the column exit cross section. The flow profiles exhibited by these columns are parabolic in shape, as was found previously for analytical and preparative size columns. The linear velocity measured in the wall region of the packing is about 2−5% lower than that in the center of the column, depending on the quality of the packed bed. The local efficiency is highest in the core region of the packed bed, where local, on-column detection would yield better separations, exhibiting a higher degree of resolution. The radial distribution of the analyte concentration is not uniform but shows important fluctuations along the radius of the column. The differences in local linear velocity, local efficiency, and local analyte concentration observed are the smallest for the most efficient columns.