Glucose sensing and hybrid instrumentation based on printed organic electrochemical transistors

Rakesh Rajendran Nair
DOI: https://doi.org/10.1088/2058-8585/ab63a0
IF: 3.1
2020-01-16
Flexible and Printed Electronics
Abstract:This report presents the design and fabrication of a fully open-air printed organic polymer biosensor for detecting hypo/hyperglycaemia among humans with a fully integrated inorganic measurement and notification circuitry fabricated on a single flexible plastic substrate. The sensor is based on the electrochemical doping and de-doping of the degenerately doped organic semiconductor Poly(3,4-ethylenedioxythiophene) poly(styrene sulphonate) (PEDOT:PSS). The printed sensor reported herein, demonstrated sensitivity between 0.001 and 10 mM glucose concentrations utilizing a PEDOT: PSS gate, hence allowing for saliva-based, non-invasive glucose measurements. Transimpedance amplification and data processing circuitry for a prototype measurement and notification system was developed using three inexpensive chips: On Semiconductor's CAT102 precision regulator, Texas Instruments' LM358DR op-amp and Microchip's-ATMEGA328PB-AUR, all integrated on the same polyethylene terephthalate substrate using screen-printed Ag interconnects and vertically conducting anisotropic glue to demonstrate the system's practicability as a cheap, self-reliant, battery powered bio-sensing module. To demonstrate the efficacy of the system 'as is', no redox mediators to amplify the biosensor response were employed.
materials science, multidisciplinary
What problem does this paper attempt to address?