Flexible Gas Sensor Printed on a Polymer Substrate for Sub-ppm Acetone Detection

W. Andrysiewicz,J. Krzeminski,K. Skarżynski,K. Marszalek,M. Sloma,A. Rydosz
DOI: https://doi.org/10.1007/s13391-020-00199-z
IF: 3.151
2020-01-20
Electronic Materials Letters
Abstract:Abstract Gas sensors are widely used in many industrial and home applications. There is therefore continued need to develop novel gas sensor substrates which provide good mechanical and electrical stability, and good flexibility in comparison with the conventional alumina and silicon-based materials. In this paper, we present the experimental results on flexible gas sensors based on the Kapton foil and alumina substrate covered by copper oxide as a gas-sensitive layer. These sensors exhibited good mechanical stability and gas-sensing characteristics. The Kapton-based CuO gas sensors were tested under exposure to acetone in the 0.05–1.25 ppm range (150 °C, 50%RH). The results confirmed that sensors deposited on the flexible substrate such as Kapton can be used in the exhaled breath analyzers dedicated to diabetes biomarker detection or other applications for which the elastic substrate is needed. Graphic Abstract
materials science, multidisciplinary
What problem does this paper attempt to address?