Evaluation of Software-Based Metal Artifact Reduction in Intraoperative 3D Imaging of the Spine Using a Mobile Cone Beam CT

Maxim Privalov,Marcus Mohr,Benedict Swartman,Nils Beisemann,Holger Keil,Jochen Franke,Paul Alfred Grützner,Sven Yves Vetter
DOI: https://doi.org/10.1007/s10278-020-00324-2
IF: 4.903
2020-07-06
Journal of Digital Imaging
Abstract:The aim of our study was to evaluate whether software-based artifact reduction can achieve an improved image quality, using intraoperative 3D imaging in spinal surgery. A total of 49 intraoperative 3D image datasets of patients, who underwent surgery with pedicle screw placement, were retrospectively evaluated. The visibility of anatomical structures and the diameter of the pedicle screws were examined, with and without the application of the artifact reduction software. All software prototypes can improve the visibility of anatomical structures (<i>P</i> &lt; 0.01), except MAR (metal artifact reduction) combined with IRIS (iterative reconstruction in image space) (<i>P</i> = 0.04). The algorithms MAR and MAR-2 can reduce the blooming artifacts significantly (<i>P</i> &lt; 0.01), but SL (Shepp &amp; Logan) cannot (<i>P</i> = 0.08–0.988). In summary, software-based artifact reduction for intraoperative 3D datasets can improve the current image quality. Additional information regarding the implant placement and the fracture reduction is therefore generated for the surgeon.
radiology, nuclear medicine & medical imaging