Evaluation of an optimized metal artifact reduction algorithm for flat-detector angiography compared to DSA imaging in follow-up after neurovascular procedures

Nadine Amelung,Volker Maus,Daniel Behme,Ismini E. Papageorgiou,Johanna Rosemarie Leyhe,Michael Knauth,Marios Nikos Psychogios
DOI: https://doi.org/10.1186/s12880-019-0352-2
IF: 2.7949
2019-08-14
BMC Medical Imaging
Abstract:<h3 class="Heading">Background</h3><p class="Para">Flat detector CT – angiography (FDCTA) has become a valuable imaging tool in post- and peri-interventional imaging after neurovascular procedures. Metal artifacts produced by radiopaque implants like clips or coils still impair image quality.</p><h3 class="Heading">Methods</h3><p class="Para">FDCTA was performed in periprocedural or follow-up imaging of 21 patients, who had received neurovascular treatment. Raw data was sent to a dedicated workstation and subsequently a metal artifact reduction algorithm (MARA) was applied. Two neuroradiologists examined the images.</p><h3 class="Heading">Results</h3><p class="Para">Application of MARA improved image appearance and led to a significant reduction of metal artifacts. After application of MARA only 8 datasets (34% of the images) were rated as having many or extensive artifacts, before MARA 15 (65%) of the images had extensive or many artifacts. Twenty percent more cases of reperfusion were diagnosed after application of MARA, congruent to the results of digital subtraction angiography (DSA) imaging. Also 3 (13% of datasets) images, which could not be evaluated before application of MARA, could be analyzed after metal artifact reduction and reperfusion could be excluded.</p><h3 class="Heading">Conclusion</h3><p class="Para">Application of MARA improved image evaluation, reduced the extent of metal artifacts, and more cases of reperfusion could be detected or excluded, congruent to DSA imaging.</p>
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?