High‐pressure stopped‐flow nuclear magnetic resonance apparatus for the study of fast reactions in solution

Shigenobu Funahashi,Koji Ishihara,Sen-ichi Aizawa,Takashi Sugata,Masao Ishii,Yasuhiro Inada,Motoharu Tanaka,Sen‐ichi Aizawa
DOI: https://doi.org/10.1063/1.1144426
IF: 1.6
1993-01-01
Review of Scientific Instruments
Abstract:A high-pressure proton nuclear magnetic resonance (NMR) probe for a stopped-flow method has been developed, which enables us to follow fast reactions at various pressures up to 200 MPa. The high-pressure stopped-flow NMR (HP-SF-NMR) method has been applied to the study of several reactions at high pressures: the reaction of tetramethyltin with iodine in chloroform and the solvent exchange reaction on aluminum(III) ion in N, N-dimethylformamide. The volumes of activation for the iodomerization of tin and the solvent exchange reaction on aluminum(III) were determined to be −24.9±3.8 cm3 mol−1 and −2.2±0.8 cm3 mol−1, respectively. The apparatus permits measurement of activation volumes for reactions with half lives longer than a few seconds if the nuclear relaxation time of the measured proton is short enough. Several requirements for the applicability of HP-SF-NMR are discussed.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?