Low-noise microwave SQUID multiplexed readout of 38 x-ray transition-edge sensor microcalorimeters

Y. Nakashima,F. Hirayama,S. Kohjiro,H. Yamamori,S. Nagasawa,A. Sato,S. Yamada,R. Hayakawa,N. Y. Yamasaki,K. Mitsuda,K. Nagayoshi,H. Akamatsu,L. Gottardi,E. Taralli,M. P. Bruijn,M. L. Ridder,J. R. Gao,J. W. A. den Herder
DOI: https://doi.org/10.1063/5.0016333
IF: 4
2020-09-21
Applied Physics Letters
Abstract:We report very-low-noise, fast-response, middle-scale multiplexing in a microwave superconducting quantum interference device multiplexer (MW-Mux) as a transition-edge sensor (TES) readout. Our MW-Mux is able to read 40 channels with 500 kHz sampling and has a low readout noise of 0.9   μ Φ 0 / Hz (where Φ 0 is the magnetic flux quantum), equivalent to 9   pA / Hz. By contrast, a multiplexer of less than 10 pixels with 500 kHz sampling and ∼ 2   μ Φ 0 / Hz readout noise has so far been reported in the literature. Owing to the 500 kHz sampling, our MW-Mux exhibits a fast response to detect a TES pulse with a rise time around 12   μ s. We demonstrated simultaneous readout of 38 pixels from an array of x-ray TES microcalorimeters. The measured full-width values at half-maximum spectral resolution ranged from 2.79 to 4.56 eV, with a median value of 3.30 eV at 5.9 keV, including a ∼ 10 % contribution of readout noise, i.e., 0.9–1.7 eV.
physics, applied