Applications for Synchrotron Radiation in Inertial Confinement Fusion Research
KX Sun,RQ Yi,TX Huang,YL Cui,JM Yang,SE Jiang,YK Ding,BH Zhang,HB Wang,SH Wen,ZJ Zheng,MQ Cui,PP Zhu,YD Zhao,G Li,ES Tang
DOI: https://doi.org/10.3321/j.issn:0254-3052.2001.z1.025
2001-01-01
Abstract:In this paper the calibrations of different soft X-ray detector elements used experiment diagnostics of ICF in Beijing Synchrotron Radiation Facility-3W1B beam is reported. SR used special operation type with beam current 35-110mA, the storage ring electron energy 2GeV, photon energy region 50-1500eV. First, property of light source was studied by transmission grating-face array CCD system. We have divided 50-1500eV energy region in seven energy section through different material prefilter with monochromaticity better 95%. The AXUV-100 photodiode produced by IRD of America as secondary standard monitored light source intensity. The successive adjust monochromatic light was contained through reflectometer target chamber and relative adjust facility. The,sensitivity of XRD(Au,Al), the reflectivity of planar mirror and multi-layer mirror, the transmissivity of different material filter, the diffraction efficiency of transmission grating, the energy response curve of soft,X-ray film, the quantum efficiency of face array CCD and the still property of soft X-ray streake camera et al. were calibrated. Finally, the experiment data treatment in Shenguang- II were used by many calibration resuite, which secure data treatment accuracy.